Viabahn endoprosthesis for femoropopliteal aneurysm repair: safety, success rates, and long-term patency

Abstract

Background

The Viabahn endoprosthesis has become a vital option for endovascular therapy, yet there is limited long-term data on its effectiveness for peripheral aneurysm repair. This study aimed to evaluate the safety, technical and clinical success, and long-term patency of the Viabahn endoprosthesis for treating femoropopliteal aneurysms.

Methods

This retrospective tertiary single-center study analyzed patients who underwent a Viabahn endoprosthesis procedure for femoropopliteal aneurysm repair from 2010 to 2020. Intraoperative complications, technical and clinical success rates, and major adverse events (MAE, including acute thrombotic occlusion, major amputation, myocardial infarction, and device- or procedure-related death) at 30 days were assessed. Incidence of clinically-driven target lesion revascularisation (cdTLR) was noted. Patency rates were evaluated by Kaplan–Meier analysis.

Results

Among 19 patients (mean age, 72 ± 12 years; 18 male, 1 female) who underwent aneurysm repair using the Viabahn endoprosthesis, there were no intraoperative adverse events, with 100% technical and clinical success rates. At the 30-day mark, all patients (19/19, 100%) were free of MAE. The median follow-up duration was 1,009 days [IQR, 462–1,466]. Popliteal stent graft occlusion occurred in 2/19 patients (10.5%) after 27 and 45 months, respectively. Consequently, the primary patency rates were 100%, 90%, 74% at 12, 24, and 36–72 months, respectively. Endovascular cdTLR was successful in both cases, resulting in sustained secondary patency at 100%.

Conclusion

The use of Viabahn endoprostheses for femoropopliteal aneurysm repair demonstrated technical and clinical success rates of 100%, a 0% 30-day MAE rate, and excellent long-term patency.

Peer Review reports

Introduction

The Viabahn endoprosthesis is a heparin-bonded stent graft that serves as an endovascular treatment option for peripheral artery disease [1,2,3], and has been expanded to include peripheral aneurysm repair [4]. Peripheral aneurysms, encompassing femoral and popliteal aneurysms, are less common than their abdominal aortic counterparts [5,6,7,8]. They typically affect men over 60 years of age and often occur alongside a variety of cardiovascular risk factors and comorbidities [9]. The potential consequences of femoropopliteal aneurysms are diverse, ranging from thrombosis and distal embolization to acute ischemia and compression of adjacent structures, with instances of rupture and limb loss [10,11,12,13,14]. In order to prevent unfavorable outcomes, it is imperative to provide optimal treatment. Elective open surgical repair via interposition or bypass graft is considered the gold standard [6, 15, 16]. Given the predominantly elderly patient demographic and the complexity of associated risks, endovascular treatment is frequently necessary due to its minimally invasive nature. The Viabahn endoprosthesis is widely recognized for its precise delivery capability and structural stability. However, despite its growing use in aneurysm therapy, there remains limited comprehensive data on its long-term performance on femoropopliteal aneurysm repair. Long-term outcome data is of significant importance in order to provide evidence-based treatment decisions, particularly in patients with a long enough life expectancy to suffer from potential long-term adverse events such as graft occlusion. Therefore, it is crucial to obtain robust performance data on safety and long-term patency, especially in a real-world setting [17, 18].

Consequently, the purpose of this investigation is to assess the safety, technical and clinical success, as well as the long-term outcomes of the Viabahn endoprosthesis for the management of femoropopliteal aneurysms.

Methods

Patient population

In this tertiary single-center study (Tuebingen, Germany), adult patients (age 18 years or older) who underwent femoropopliteal aneurysm repair using the Viabahn endoprosthesis between July 2010 and October 2020 were consecutively recruited and retrospectively analyzed. Endovascular treatment was indicated following the decision of an interdisciplinary tumor board for symptomatic aneurysms or for aneurysms with a diameter of ≥ 25 mm in femoral aneurysms and ≥ 20 mm in popliteal aneurysms [6, 16]. An all-comers approach was used. Cardiovascular risk factors were comprehensively documented, encompassing dyslipidemia, family history of premature coronary artery disease, hypertension, non-insulin-dependent (NIDDM) or insulin-dependent (IDDM) diabetes mellitus, obesity, and smoking status categorized as current, former, or never smoked. Comorbidities were recorded, renal function was assessed by measuring the glomerular filtration rate (GFR), and chronic renal failure (GFR < 60 mL/min/1.73m2) or current hemodialysis were noted. Evidence of bacteremia and known uncorrectable hypercoagulability were documented. The study received approval from the institutional review board (approval number: 254/2021BO1, date: 2021/08/20) and complied with the regulations of the Health Insurance Portability and Accountability Act (HIPAA), the need for informed consent was waived.

Study procedure

All procedures were conducted under local anesthesia or conscious sedation. Following the establishment of vascular access, an angiographic catheter was navigated to the distal aspect of the target artery with the assistance of a guidewire. Subsequently, angiography of the target artery was conducted for aneurysm identification. The guidewire was replaced with a stiff wire, and the sheath was substituted with a dedicated 6F to 12F sheath. The selection of stent graft size was tailored to each patient based on pre-procedural imaging and/or intraoperative measurements. The diameter and length of the stent graft were chosen to ensure optimal apposition to the vessel wall and adequate coverage of the aneurysm. The sheath was advanced either to the distal side or as close as possible to the intended target artery. To deliver the stent graft, a stiff 0.018- or 0.035-inch wire was used. The Viabahn endoprosthesis, with diameters ranging from 5 to 13 mm and lengths ranging from 2.5 to 25.0 cm, was inserted and positioned over the intended target artery site for deployment. The characteristics of the implantation procedure were evaluated, including the number of stents employed and the dimensions (diameter in mm and length in cm) of the endoprostheses. Following the deployment of the device, immediate angiography was conducted to confirm the complete elimination of the aneurysm. Heparin (5,000 IU) was administered for the purpose of anticoagulation. The vascular access site was subsequently closed using a dedicated vascular closure device (for sheaths ≥ 6F). For sheaths exceeding 8F, suture-mediated closure devices were used for preclosing. Following the procedure, anticoagulation therapy was advised, with the recommendation of dual antiplatelet therapy, comprising daily doses of aspirin (100 mg) and clopidogrel (75 mg), to be administered for a minimum of six months, in order to reduce the risk of stent graft thrombosis. Consequently, a single dose of aspirin (100 mg per day) was prescribed as lifelong antiplatelet therapy.

Technical and clinical success rates and follow-up

The acute technical success rate was assessed according to predetermined criteria, which included the successful placement of the Viabahn stent graft, the complete elimination of perfusion of the aneurysm sac, and the preservation of blood flow in the artery, as verified by final target site angiography. The clinical success criterion was the sustained blood flow through the repaired vessel for 30 days. Prior to hospital discharge, all patients underwent a CT angiography or ultrasound evaluation to confirm adequate blood flow. A follow-up care protocol was advised subsequent to the index procedure. Physical examination and duplex ultrasound visits were scheduled at regular 30-day, 3-month, 6-month, and 12-month intervals, with annual check-ups thereafter.

Study endpoints

The primary safety endpoint of the study was the occurrence of major adverse events (MAE) within 30 days, including acute thrombotic occlusion, myocardial infarction, major limb amputation, and death. The secondary efficacy endpoint comprised the primary and secondary patency rates. In the event of stenosis or occlusion of the endoprosthesis, the time point of diagnosis after the initial procedure was documented, as well as the type of clinically driven target lesion revascularization (cdTLR) procedure, whether surgical or endovascular. The incidence of major or minor amputation was recorded.

Statistical analysis

The normality of data was assessed by the Shapiro–Wilk test. For parametric data, continuous variables are expressed as the mean ± standard deviation, whereas non-parametric data are presented as the median [IQR]. Categorical data are provided as the numerator/denominator (percentage). The Kaplan–Meier method was used to estimate patency survival curves (SPSS version 29.0, IBM).

Results

Patient characteristics

A total of 19 patients underwent endovascular aneurysm repair utilizing the Viabahn endoprosthesis, comprising 18 men and 1 woman, with a mean age of 72 ± 12 years. A detailed account of the baseline characteristics of the patients is provided in Table 1. Arterial hypertension was the most prevalent cardiovascular risk factor, detected in 11 out of 19 patients (57.9%), followed by dyslipidemia (8/19, 42.1%), and current smoking (7/19, 36.8%). The most prevalent comorbidity was coronary artery disease, present in 7 out of 19 patients (36.8%), followed by chronic obstructive pulmonary disease in 4 out of 19 patients (21.1%). The mean GFR was 74 ± 25 mL/min/1.73m2. Chronic renal failure was observed in 4 out of 19 patients (21.1%), and none of the patients were undergoing hemodialysis. Notably, none of the patients tested positive for bacteremia, nor exhibited uncorrectable hypercoagulability.

Table 1 Patient characteristics

Procedural characteristics

Aneurysms were targeted in the popliteal artery in 16 out of 19 patients (84.2%), and in the superficial femoral artery in 3 out of 19 patients (15.8%), see Figs. 1, 2, 3 for exemplary cases. Regarding aneurysm morphology, 16/19 cases (84.2%) were fusiform aneurysms, 2/19 cases (10.5%) were saccular aneurysms, and 1/19 cases (5.3%) was an eccentric aneurysm. Aneurysms with thrombus occupying more than 50% of the lumen were recorded in 9/19 patients (47.4%). A tibial runoff of less than 2 vessels was observed in 5 out of the 19 patients (26.3%). A total of 26 endoprostheses were placed, with the most frequently employed device diameter measuring 6 mm (13/26, 50.0%), followed by > 8 mm (7/26, 26.9%), 8 mm and 7 mm (each 3/26, 11.5%), as shown in Table 2. The most commonly utilized stent graft length was 10.0 cm (10/26, 38.5%), followed by 15.0 cm and 5.0 cm (each 5/26, 19.2%). Devices measuring 25.0 cm were utilized in four out of 26 cases (15.4%), 7.5 cm and 2.5 cm were used in each one out of 26 cases (3.8%). There were no intraoperative adverse events, and technical success was achieved in all 19/19 procedures (100%).

Fig. 1
figure 1

Viabahn endoprosthesis for endovascular treatment of popliteal aneurysm. Example of a 61-year-old man with peripheral artery disease and moderate claudication. A CT angiography (CTA) revealed a popliteal aneurysm and stent fracture in the P2 segment. B Subsequently, the patient underwent Viabahn stent graft procedure. Final target site angiography demonstrated successful aneurysm exclusion by the implanted Viabahn with preserved blood flow and no peripheral complications. C Follow-up CTA at 2 years demonstrated sustained blood flow through the device

Fig. 2
figure 2

Viabahn endoprosthesis for popliteal aneurysm repair. A 73-year-old man was referred for (A) CT angiography, which revealed a popliteal aneurysm (arrows) with substantial lumen thrombus (stars). B Stent graft repair was performed using a Viabahn endoprosthesis (proximal and distal ends of the stent graft indicated by arrowheads). C Final target site angiography confirmed successful exclusion of the aneurysm and preservation of blood flow

Fig. 3
figure 3

Viabahn endoprosthesis for femoral aneurysm repair. A 77-year-old man with previous P1/P3 bypass operation presented with swelling and pain in the distal posterior aspect of his thigh. A CT angiography (CTA) revealed the presence of a false aneurysm of the distal superficial femoral artery (arrow). B The patient underwent stent graft repair using a Viabahn endoprosthesis (indicated by arrowheads). The successful exclusion of the aneurysm with preservation of blood flow was confirmed by final target site angiography. C Follow-up CTA at 1 year demonstrated sustained preservation of blood flow

Table 2 Procedural characteristics

Clinical success and major adverse events within 30 days

All 19/19 procedures (100%) demonstrated clinical success, as shown in Table 3. There were no device- or procedure-related deaths, myocardial infarctions, major amputations, or acute thrombotic occlusions within 30 days of stent graft procedure (0/19 patients, 0.0%). Thus, all patients (19/19, 100%) were free from any major adverse events at the 30-day mark.

Table 3 Clinical and procedural outcomes

Long-term outcome

The median follow-up duration was 1,009 days [IQR, 462–1,466], with a range of 34 to 2,806 days. Two cases of popliteal stent graft occlusion were observed. One graft occlusion of a 6 mm Viabahn device occurred at 27 months post-procedure in a 58-year-old male patient with arterial hypertension, dyslipidemia, current smoking, and a family history of premature coronary artery disease. Another graft occlusion of a 10 mm Viabahn device occurred at 45 months post-procedure in a 75-year-old male patient with non-insulin dependent diabetes mellitus, chronic renal failure, a family history of premature coronary artery disease, and a two-vessel run-off. In both instances, endovascular revascularization was successful. The patient with popliteal occlusion at 27 months had pre-existing stage IV (Fontaine) peripheral artery disease, manifested by poor distal run-off (collateral vessels only). Despite the successful endovascular revascularization of the stent graft, lower leg amputation was subsequently considered the ultimate treatment option. Kaplan–Meier analysis unveiled primary patency rates of 100%, 90%, and 74% at 12, 24, and 36–72 months post-endoprosthesis procedure, respectively, see Fig. 4. Secondary patency was observed to be sustained at 100%.

Fig. 4
figure 4

Long-term primary patency of the Viabahn endoprosthesis for femoropopliteal aneurysm management. The curve shows the primary patency (%) of the Viabahn endoprosthesis for the treatment of femoropopliteal aneurysms over a period of 72 months following the procedure

Discussion

This study comprehensively evaluated the performance of the Viabahn endoprosthesis for femoropopliteal aneurysm management. The device demonstrated a good safety and success profile, with no intraoperative adverse events, technical and clinical success rates reaching 100%, and a 30-day major adverse event rate of 0%. Long-term outcomes were favorable, with primary patency rates of 100%, 90%, and 74% at 12, 24, and 36–72 months, respectively. Secondary patency was sustained at 100%.

The high technical success rate can be attributed to the excellent maneuverability facilitated by the high flexibility of the device when navigating complex arterial structures. The absence of intraoperative complications further substantiates the safety profile of the device. The 30-day major adverse event rate of 0% observed in our study provides valuable evidence supporting the device's safety across various pathologies, including stenosis, vessel injury, and aneurysms [16, 19,20,21,22,23,24]. Our results also showed sustained efficacy in the long-term, with primary patency rates of 100%, 90%, and 74% at 12, 24, and 36–72 months post initial implantation, respectively. The gradual modest decrease in primary patency rates over the long-term is consistent with previously reported reported primary patency rates at 60 months of 69%, confirming the long-term durability of Viabahn stents in the treatment of peripheral aneurysms [4]. This may be attributed, in part, to the device's capacity to withstand compressional stress during limb movement without fracture of the prosthesis, which is of critical importance for ensuring adequate perfusion distally [25]. In accordance with previous studies, we observed a high secondary patency, thereby reinforcing the concept that the Viabahn is an effective conduit for revascularization procedures [4, 26]. However, several factors can influence the patency of stent grafts, including cardiovascular risk factors, the size of the stent graft, the effectiveness of anticoagulation therapy, the presence of peripheral artery disease, and distal runoff. Each of these factors plays a critical role in determining the long-term success of the procedure and should be carefully considered during patient evaluation and treatment planning. In both instances of stent graft occlusion observed in our study, the patients exhibited cardiovascular risk factors and impaired distal run-off. Additionally, one patient had a smaller device diameter of 6 mm, which had previously been demonstrated to be associated with a reduced primary patency [27]. The findings of our study, including technical and clinical success as well as patency, demonstrate the efficacy of endovascular strategies, particularly for patient populations with underlying health conditions or complex anatomy requiring minimally invasive treatment. The advantages of endovascular aneurysm repair include a lower incidence of wound complications and a shorter hospital stay compared to open surgical treatment [15, 16]. Nevertheless, open repair remains an option in aneurysm therapy, shown to carry a lower risk of primary patency loss within the initial three years post-procedure; however, secondary patency rates are comparable between open and endovascular treatments for popliteal aneurysm treatment [15, 16]. With regard to the inherent limitations of open surgery [28], including the risk of access-related injuries and prolonged healing, the endovascular aneurysm management represents a viable alternative, offering durable arterial repair [4, 16]. Taken together, choosing the most appropriate approach requires a comprehensive assessment of lesion characteristics, patient-related factors, and treatment objectives to minimize procedural risks. This requires adopting a collaborative, interdisciplinary framework for decision-making that incorporates the expertise of both open surgical and interventional specialists, ultimately aiming to advance patient care.

This study has certain limitations that warrant consideration. First, it employs a retrospective, single-arm analysis design and lacks a control group, thereby limiting the ability to draw direct comparative conclusions. Second, the examination is limited to a patient population from a single tertiary center, which constrains the generalizability of our findings to other centers and countries. Third, the selected real-world approach, while inclusive, may pose challenges in the identification and management of confounding variables.

Conclusions

The use of the Viabahn endoprosthesis for femoropopliteal aneurysm repair had 100% technical and clinical success rates, a 0% major adverse event rate within 30 days, and yielded excellent long-term primary and secondary patency. Further studies are needed to fully evaluate the factors that influence long-term performance in order to further refine the profiling of patients who derive optimal benefits from endovascular repair over open surgical repair.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

cdTLR:

Clinically driven target lesion revascularization

IDDM:

Insulin-dependent diabetes mellitus

MAE:

Major adverse events

References

  1. Lammer J, Zeller T, Hausegger KA, Schaefer PJ, Gschwendtner M, Mueller-Huelsbeck S, et al. Heparin-bonded covered stents versus bare-metal stents for complex femoropopliteal artery lesions: The randomized VIASTAR trial (viabahn endoprosthesis with propaten bioactive surface [VIA] versus bare nitinol stent in the treatment of long lesions in sup. J Am Coll Cardiol. 2013;62(15):1320–7. https://doi.org/10.1016/j.jacc.2013.05.079.

    Article  CAS  PubMed  Google Scholar 

  2. Geraghty PJ, Mewissen MW, Jaff MR, Ansel GM. Three-year results of the VIBRANT trial of VIABAHN endoprosthesis versus bare nitinol stent implantation for complex superficial femoral artery occlusive disease. J Vasc Surg. 2013;58(2):386-395.e4. https://doi.org/10.1016/j.jvs.2013.01.050.

    Article  PubMed  Google Scholar 

  3. Aboyans V, Ricco JB, Bartelink MLEL, Björck M, Brodmann M, Cohnert T, et al. 2017 ESC Guidelines on the diagnosis and treatment of peripheral arterial diseases, in collaboration with the European Society for Vascular Surgery (ESVS). Eur Heart J. 2018;39(9):763–816. https://doi.org/10.1093/eurheartj/ehx095.

    Article  PubMed  Google Scholar 

  4. Patel SR, Hughes CO, Jones KG, Holt PJE, Thompson MM, Hinchliffe RJ, et al. A systematic review and meta-analysis of endovascular popliteal aneurysm repair using the Hemobahn/Viabahn stent-graft. J Endovasc Ther. 2015;22(3):330–7. https://doi.org/10.1177/1526602815579252.

    Article  PubMed  Google Scholar 

  5. Trickett JP, Scott RAP, Tilney HS. Screening and management of asymptomatic popliteal aneurysms. J Med Screen. 2002;9(2):92–3. https://doi.org/10.1136/jms.9.2.92.

    Article  CAS  PubMed  Google Scholar 

  6. Piffaretti G, Mariscalco G, Tozzi M, Rivolta N, Annoni M, Castelli P. Twenty-year experience of femoral artery aneurysms. J Vasc Surg. 2011;53(5):1230–6. https://doi.org/10.1016/j.jvs.2010.10.130.

    Article  PubMed  Google Scholar 

  7. Rana MA, Kalra M, Oderich GS, De Grandis E, Gloviczki P, Duncan AA, et al. Outcomes of open and endovascular repair for ruptured and nonruptured internal iliac artery aneurysms. J Vasc Surg. 2014;59(3):634–44. https://doi.org/10.1016/j.jvs.2013.09.060.

    Article  PubMed  Google Scholar 

  8. Brunkwall J, Hauksson H, Bengtsson H, Bergqvist D, Takolander R, Bergentz SE. Solitary aneurysms of the iliac arterial system: an estimate of their frequency of occurrence. J Vasc Surg. 1989;10(4):381–4. https://doi.org/10.1067/mva.1989.13733.

    Article  CAS  PubMed  Google Scholar 

  9. Diwan A, Sarkar R, Stanley JC, Zelenock GB, Wakefield TW. Incidence of femoral and popliteal artery aneurysms in patients with abdominal aortic aneurysms. J Vasc Surg. 2000;31(5):863–9. https://doi.org/10.1067/mva.2000.105955.

    Article  CAS  PubMed  Google Scholar 

  10. Cervin A, Acosta S, Hultgren R, Grip O, Björck M, Falkenberg M. Results After Open and Endovascular Repair of Popliteal Aneurysm: A Matched Comparison Within a Population Based Cohort. Eur J Vasc Endovasc Surg. 2021;61(6):988–97. https://doi.org/10.1016/j.ejvs.2021.02.007.

    Article  PubMed  Google Scholar 

  11. Connor D, Sharp M, Rajagopalan S. Profunda femoris artery aneurysm causing local deep venous thrombosis. J Vasc Surg. 2013;57(5):1402. https://doi.org/10.1016/j.jvs.2012.01.071.

    Article  PubMed  Google Scholar 

  12. Harbuzariu C, Duncan AA, Bower TC, Kalra M, Gloviczki P. Profunda femoris artery aneurysms: Association with aneurysmal disease and limb ischemia. J Vasc Surg. 2008;47(1):31–5. https://doi.org/10.1016/j.jvs.2007.09.040.

    Article  PubMed  Google Scholar 

  13. Dawson I, Sie RB, van Bockel JH. Atherosclerotic popliteal aneurysm. Br J Surg. 1997;84(3):293–9.

    CAS  PubMed  Google Scholar 

  14. Raine NM, Magee TR, Galland RB. Thigh embolisation in association with bilateral profunda femoris aneurysms. Eur J Vasc Endovasc Surg Off J Eur Soc Vasc Surg. 1995;9(4):491–3. https://doi.org/10.1016/s1078-5884(05)80024-3.

    Article  CAS  Google Scholar 

  15. Shahin Y, Barakat H, Shrivastava V. Endovascular versus open repair of asymptomatic popliteal artery aneurysms: a systematic review and meta-analysis. J Vasc Interv Radiol. 2016;27(5):715–22. https://doi.org/10.1016/j.jvir.2016.02.026.

    Article  PubMed  Google Scholar 

  16. Leake AE, Segal MA, Chaer RA, Eslami MH, Al-Khoury G, Makaroun MS, et al. Meta-analysis of open and endovascular repair of popliteal artery aneurysms. J Vasc Surg. 2017;65(1):246-256.e2. https://doi.org/10.1016/j.jvs.2016.09.029.

    Article  PubMed  Google Scholar 

  17. Dake MD, Ansel GM, Jaff MR, Ohki T, Saxon RR, Smouse HB, et al. Durable clinical effectiveness with paclitaxel-eluting stents in the femoropopliteal artery: 5-year results of the Zilver PTX randomized trial. Circulation. 2016;133(15):1472–83. https://doi.org/10.1161/CIRCULATIONAHA.115.016900.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Iida O, Urasawa K, Komura Y, Soga Y, Inoue N, Hara H, et al. Self-expanding nitinol stent vs percutaneous transluminal angioplasty in the treatment of femoropopliteal lesions: 3-year data from the SM-01 Trial. J Endovasc Ther. 2019;26(2):158–67. https://doi.org/10.1177/1526602819826591.

    Article  PubMed  Google Scholar 

  19. Mafeld S, Annamalai G, Lindsay TF, Zhong I, Tarulli E, Mironov O, et al. Initial experience with Viabahn VBX as the bridging stent graft for branched and fenestrated endovascular aneurysm repair. Vasc Endovascular Surg. 2019;53(5):395–400. https://doi.org/10.1177/1538574419840880.

    Article  PubMed  Google Scholar 

  20. Bismuth J, Gray BH, Holden A, Metzger C, Panneton J. Pivotal study of a next-generation balloon-expandable stent-graft for treatment of iliac occlusive disease. J Endovasc Ther. 2017;24(5):629–37. https://doi.org/10.1177/1526602817720463.

    Article  PubMed  Google Scholar 

  21. Segal A, Flugelman MY, Khader N, Rubinshtein R, Lavi I, Karmeli R, et al. Outcome of stent graft implantation for treatment of access site bleeding after transfemoral transcatheter aortic valve replacement. Am J Cardiol. 2017;120(3):456–60. https://doi.org/10.1016/j.amjcard.2017.04.050.

    Article  PubMed  Google Scholar 

  22. Ueda T, Murata S, Tajima H, Saito H, Yasui D, Sugihara F, et al. Emergency endovascular treatment using a Viabahn stent graft for upper and lower extremity arterial bleeding: a retrospective study. CVIR Endovasc. 2021;4(1). https://doi.org/10.1186/s42155-021-00273-z.

  23. Schahab N, Kavsur R, Mahn T, Schaefer C, Kania A, Fimmers R, et al. Endovascular management of femoral accesssite and access-related vascular complications following percutaneous coronary interventions (PCI). PLoS ONE. 2020;15(3):1–12. https://doi.org/10.1371/journal.pone.0230535.

    Article  CAS  Google Scholar 

  24. Brendel JM, Mangold T, Lescan M, Schmehl J, Ghibes P, Grimm A, et al. Viabahn stent graft for arterial injury management: safety, technical success, and long-term outcome. CVIR Endovasc. 2024;7(1). https://doi.org/10.1186/s42155-024-00435-9.

  25. Lee Y-J, Shin D-H, Kim J-S, Kim B-K, Ko Y-G, Hong M-K, et al. Femoropopliteal artery stent fracture with recurrent in-stent reocclusion and aneurysm formation: successful treatment with self-expandable viabahn endoprosthesis. Korean Circ J. 2015;45(6):522–5. https://doi.org/10.4070/kcj.2015.45.6.522.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Guzzardi G, Natrella M, Del Sette B, Petullà M, Fanelli G, Porta C, et al. Endovascular repair of popliteal artery aneurysms: an Italian multicenter study. Radiol Medica. 2019;124(1):79–85. https://doi.org/10.1007/s11547-018-0941-0.

    Article  Google Scholar 

  27. Squizzato F, Mosquera-Rey V, Zanabili Al-Sibbai A, Camblor Santervas LA, Pasqui E, Palasciano G, et al. Outcomes of Self-Expanding Covered Stents for the Treatment of External ILIAC Artery Obstructive Disease. Cardiovasc Intervent Radiol. 2023;46(5):579–87. https://doi.org/10.1007/s00270-023-03370-9.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Lawrence PF, Harlander-Locke MP, Oderich GS, Humphries MD, Landry GJ, Ballard JL, et al. The current management of isolated degenerative femoral artery aneurysms is too aggressive for their natural history. J Vasc Surg. 2014;59(2):343–9. https://doi.org/10.1016/j.jvs.2013.08.090.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was not supported by any funding.

Author information

Authors and Affiliations

Authors

Contributions

J.M.B.: Conceptualization, Methodology, Investigation, Data curation, Formal Analysis, Writing—Original draft, Visualization, Project administration. T.M.: Data curation, Writing—review & editing. M.P.: Data curation, Writing—review & editing. B.S.: Writing—review & editing. M.A.: Writing—review & editing. J.M.: Writing—review & editing. J.S.: Writing—review & editing. P.K.: Writing—review & editing. C.A.: Conceptualization, Methodology, Supervision, Writing—review & editing. G.G.: Conceptualization, Methodology, Supervision, Writing—review & editing, Project administration. A.E.: Conceptualization, Methodology, Supervision, Writing—review & editing, Project administration. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Gerd Grözinger.

Ethics declarations

Ethical approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The study has obtained approval form the institutional review board (IRB approval number: 254/2021BO1, date: 2021/08/20) and the need for informed consent was waived.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Brendel, J.M., Mangold, T., Pfändler, M. et al. Viabahn endoprosthesis for femoropopliteal aneurysm repair: safety, success rates, and long-term patency. CVIR Endovasc 7, 53 (2024). https://doi.org/10.1186/s42155-024-00465-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42155-024-00465-3

Keywords