Table 1 Exposure-parameters for phantom study

From: Evaluation of novel radiation protection devices during radiologically guided interventions

Parameters

Head protector and ceiling mounted lead shield

Extended thyroid protector and standard thyroid collar

Modality

Siemens Artis Qa

Philips Multi Diagnost Eleva FDb

Table height (cm)

90

80

Operator phantom height, h (cm)

165

165

Central beam axis—operator phantom front distance, d (cm)

26

25

Central beam axis—operator phantom side distance, s (cm)

58

66

Protocol

Abdomen

Abdomen

Exposure mode

Acquisition

Acquisition

Zoom (cm)

42

48

Source image distance

120

125

Patient phantom surface—detector distance, y (cm)

50

54

Frames per second

4

2

Tube voltage (kV)

70

90

Tube load (mAs)

30

25

Filter

No

1 mm Al, 0.1 mm Cu

DAPc (Gycm2)

400

130

  1. aSiemens Artis Q (Siemens Healthineers, Germany) angio system
  2. bPhilips MultiDiagnost Eleva FD (Philips, The Netherlands)
  3. cDose Area Product, to terminate the radiation at the same exposure-level