Table 6 Comparison of procedure number, mean fluoroscopy time, air kerma-area product, reference air kerma between the RAD-IR study, old fluoroscopy system and new fluoroscopy system

From: Patient radiation dose during angiography and embolization for abdominal hemorrhage: the influence of CT angiography, fluoroscopy system, patient and procedural variables

 

N

FT (min)

 

PKA (cGy.cm2)

 

Ka,r (mGy)

 
 

Mean (95% CI)

Range

Mean (95% CI)

Range

Mean (95% CI)

Range

RAD-IR

94

25.8 (22.2–29.5)

3.5–93.7

34,757 (30,599–38,915)

2713– 129,465

2367 (2037–2697)

105–7160

OS

61

19.6 (16.4–22.8)

1.33–61.9

28,540 (23,524-33,555)

138.5–96,876

1688 (1074-2302)

100–17,100

% Reduction compared to RAD-IR

 

24.0

 

17.9

 

28.7

 

NS

93

20.8 (18.1–23.4)

0.6–62.4

11, 875 (8979-14,770)

261–63,118

778 (594–962)

52.8–3323

% Reduction compared to RAD-IR

 

19.4

 

65.8

 

67.1

 
  1. Note: The RAD-IR study used mean and not median values for FT, PKA and Ka,r. Percentage reduction of PKA and Ka,r are calculated relative to the RAD-IR mean dose. N number, OS old system (Axiom Artis DTA), NS new system (Axiom Artis Q), BMI body mass index, FT fluoroscopy time, PKA air kerma-area product, Ka,r reference air kerma. N/A not available