Table 5 Univariate and multivariate linear regression modelling of factors influencing radiation dose (n = 154)

From: Patient radiation dose during angiography and embolization for abdominal hemorrhage: the influence of CT angiography, fluoroscopy system, patient and procedural variables

 

Univariate

Multivariate

R2

B

95% CI

p

R2

B

95% CI

p

Air-Kerma Product (PKA)

  

0. 459

  

< 0.005

 Age

0.005

77.52

(−96)-251

0.379

    

 Duration

0.058

124

44–204

0.003

 

112

(−5.6)-231

0.062

 FT

0.046

5.2

1.42–8.99

0.007

 

0.229

(−5.07)-5.53

0.932

 DSA runs

0.158

1638

1028–2248

< 0.005

 

929

309–1548

0.004

 Location

0.011

− 2307

(− 6306)-2050

0.197

    

 Indication

0.004

− 1752

(− 5766)-1692

0.449

    

 BMI

0.117

5701

3394–7970

0.001

 

5714

3632–7795

< 0.005

 FS

0.199

−16,665

(−22,438)- (−10,800)

0.001

 

16,561

21,258–11,863

< 0.005

Reference Air Kerma (Ka,r)

  

0.152

  

< 0.005

 Age

0.005

6.8

(−10)-24

0.426

    

 Duration

0.013

5.7

(−2.63)-14.02

0.179

    

 FT

0.012

0.255

(−0.142)-0.653

0.206

    

 DSA runs

0.066

97.68

35.18–160.19

0.002

 

87

26.65–147.33

0.005

 Location

0.006

− 171

(− 533)-189

0.329

    

 Indication

0.002

−117

(− 552)-317

0.595

    

 BMI

0.034

280

26–535

0.031

 

240

(−3.31)-483

0.053

 FS

0.071

909

350–1467

0.002

 

822

1364–280

0.003

  1. Table 5 displays the results of both the univariate and multivariate regression analysis of the patient and procedural factors influence on patient radiation dose, measured as both PKA (air kerma-area product) and Ka,r (reference air kerma). Only those factors that were found to have a significant impact on radiation dose during the univariate analysis were included in the multivariate analysis
  2. FT Fluoroscopy Time, FS Fluoroscopy System