Skip to main content

Table 3 Filter feature assessed on angioCT-scan

From: Safety and efficacy of the VenaTech™ Retrievable inferior vena cava filter: a first-in-man single-center prospective study

Data

Temporary population

Permanent population

Number

21

10

Delay from implantation (days)

42 ± 25

146 ± 71

IVC occlusion

0

0

Filter thrombosis/embolization:

3 (14)

0

Filter fracture

0

0

Filter migration

0

0

Tilt:

  < 15°

2 (9)

2 (20)

  > 15°

3 (14)

2 (20)

IVC wall penetration:

  < 3 mm

4 (19)

2 (20)

  > 3 mma

1 (5)

3 (30)

  1. Values are provided as Mean ± SD (range) or n (%)
  2. aIVC wall penetration were respectively 4 mm in the temporary population and, 4, 5 and 5 mm in the permanent population and were limited to fat surrounding the IVC