Table 2 Procedural data at implantation and retrieval

From: Safety and efficacy of the VenaTech™ Retrievable inferior vena cava filter: a first-in-man single-center prospective study

Data

Value

Level of deployment at insertion:

 T12

1 (2)

 L1

10 (25)

 L1/L2

2 (5)

 L2

17 (42)

 L2/L3

1 (2)

 L3

6 (15)

 L4

3 (7)

Anticoagulation regimen at retrieval:

 None

1 (5)

 Direct oral anticoagulants

1 (5)

 Low-Molecular-Weight Heparins

19 (90)

Reasons for moderate difficulty to retrieve:

 Anterior tilt

1 (5)

 Tilt > 15°

2 (9)

 IVC thrombus

1 (5)

Additional retrieval devices:

 Grasping device (ALN® kit)

3 (14)

 10 Fr. manual aspiration catheter

1 (5)

Reasons for failure of retrieval:

 Posterior tilt

1 (5)

 Filter thrombosis

2 (9)

 Time to retrieval failure (days)

[68, 70, 72]

  1. Values are provided as n(%) or [exact value]