Skip to main content

Table 1 Aneurysm type, Risk factors and Modification specifications

From: Clinical application and technical details of cook zenith devices modification to treat urgent and elective complex aortic aneurysms

Subject Aneurysm type Aneurysm size (cm) Cardiovascular risk factors (CVRFs) Device used Number of fenestrations Modification time (minutes)
1 JRAA 8.5 Hypertension
Hyperlipidemia
COPD
Coronary artery disease
Tobacco abuse
ASA score: 4
Zenith Flex AAA Endovascular Graft Bifurcated Main Body 3
- SMA
- RRA
- LRA
98
2 TAAA 7.6 Hypertension
Hyperlipidemia
Obesity
COPD
Aortic Dissection
ASA score: 4
Zenith TX2 Dissection Endograft with Pro-Form 5
- CA
- SMA
- RRAx2
- LRA
128
3 Failed previous EVAR 10.6 Hypertension
Hyperlipidemia
COPD
CKD stage III
Coronary artery disease
Tobacco abuse
ASA score: 4
Zenith Alpha Thoracic Stent Graft 3
- SMA
- RRA
- LRA
103
4 TAAA, extent V 6.8 Hypertension
Hyperlipidemia
COPD
Coronary artery disease
Tobacco abuse
ASA score: 4
Cook Fenestrated Stent Graft (Zfen) with 2 existing fenestrations for CA and SMA 4
- CA
- SMA
- RRA
- LRA
27
  1. JRAA: Juxtarenal aneurysm; TAAA: Thoracoabdominal aortic aneurysm; COPD: chronic obstructive pulmonary disease; CKD III: chronic kidney disease, stage III; ASA: American Society of Anesthesiologists; SMA: superior mesenteric artery; RRA: right renal artery; LRA: left renal artery; CA: celiac axis