Table 3 Quantitative In Vivo Velocities

From: A technique for intra-procedural blood velocity quantitation using time-resolved 2D digital subtraction angiography

  MR Velocity (cm/s) DSA Velocity (cm/s) DSA Std. Dev. Mag. Factor
Swine 1
 Common Hepatic 53.51 53.53 6.79 1.19
 Splenic 27.15 26.33 4.53 1.11
 Left Renal 22.57 16.65 1.98 1.28
 Right Renal 25.88 31.44 4.39 1.16
 Iliac 38.33 33.28 5.03 1.20
Swine 2
 Common Hepatic 43.01 47.78 8.78 1.28
 Splenic 29.56 34.16 2.72 1.26
 Right Renal 29.62 35.21 4.48 1.28
 Iliac 34.05 26.38 3.11 1.33
  1. Quantitative values from the in vivo study for both imaging modalities are presented in the table. The standard deviation of the DSA velocity is provided from the triplicate scans. The magnification factor was calculated for each vessel by imaging a reference wire, of 3 cm in length, in the vessel of interest prior to injection