Skip to main content

Table 2 Procedural metrics of 121 patients undergoing fenestrated endovascular aortic repair (FEVAR) using the Zenith Fenestrated AAA Endovascular® with large fenestrations for superior mesenteric artery incorporation

From: Superior mesenteric artery outcomes after large fenestration strut relocation with the Zenith Fenestrated endoprosthesis

 

Total

(n = 121)

No relocation

(n = 27)

Relocation

(n = 94)

P value

Median (Q1, Q3)

Total operative time (min)

88 (77, 157)

210 (163, 250)

80 (72, 90)

< 0.001

Fluoroscopy time (min)

52 (46, 66)

80 (61, 94)

51 (45, 61)

< 0.001

Radiation dose (mGy)

1098 (989, 1670)

1609 (1105, 1982)

1090 (978, 1312)

< 0.001

Contrast volume (mL)

52 (37, 64)

67 (60, 90)

47 (35, 57)

< 0.001

  1. Min Minutes, mGy milliGray, mL Milliliters