From: Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting

Fig. 2

Aspiration volume flow rates at a pressure of 50 mmHg (y-axis, mL = Milliliters) plotted vs remaining axial surface area of the guide catheter working lumen after the stent catheter was introduced (x-axis, mm2 = Millimeters square). Plotted are aspiration rates with all undeployed stents and the stent catheter shafts of the models where size information was available within the working lumen of the guide catheter (Abbreviations: Acc: Acculink, Prec 7/10: Precise 7 mm/10 mm, Prot: Protégé. The numbers in brackets show the volumetric aspiration rates). The graph depicts a nearly linear relation of flow rates and luminal surface area, which shows that laminar flow is not present