Skip to main content

Table 5 Major vascular closure device / puncture site complications during learning and established procedure

From: Access site complications after transfemoral aortic valve implantation - a comparison of Manta and ProGlide

  ProGlide learning (n = 25) Manta learning (n = 25) p ProGlide established (n = 51) Manta established (n = 50) p
All complications (occlusion, bleeding, dissection, stenosis or pseudo aneurysm), n (%) 1 (4.0) 2 (8.0) 0.551 1 (2.0) 6 (12.0) 0.047
Occlusion, n (%) 1 (4.0) 0 (0) 0 (0) 1 (2.0)
Bleeding, n (%) 1 (4.0) 1 (4.0) 1 (2.0) 5 (10.0)
Pseudo aneurysm, n (%) 0 (0) 2 (8.0) 0 (0) 2 (4.0)
Flow limiting stenosis or dissection, n (%) 1 (4.0) 0 (0) 0 (0) 1 (2.0)